恒(heng)(heng)溫恒(heng)(heng)濕(shi)試(shi)(shi)(shi)驗(yan)(yan)(yan)箱控制(zhi)方式(shi)與特殊:平衡調(diao)溫調(diao)濕(shi)控制(zhi)系(xi)統(tong)(BTHC),以P.I.D.方式(shi)控制(zhi)SSR,使(shi)系(xi)統(tong)之加熱(re)加濕(shi)量(liang)(liang)等于熱(re)濕(shi)損(sun)耗量(liang)(liang),故能長期穩定的使(shi)用。但是(shi)恒(heng)(heng)溫恒(heng)(heng)濕(shi)試(shi)(shi)(shi)驗(yan)(yan)(yan)設(she)備而是(shi)能夠(gou)滿(man)足(zu)自(zi)己實驗(yan)(yan)(yan)條件(jian)(jian),滿(man)足(zu)產(chan)(chan)品(pin)(pin)實驗(yan)(yan)(yan)標準,首先將被(bei)試(shi)(shi)(shi)驗(yan)(yan)(yan)產(chan)(chan)品(pin)(pin)(元器件(jian)(jian)、組件(jian)(jian)、部件(jian)(jian)或整(zheng)機)置(zhi)入恒(heng)(heng)溫恒(heng)(heng)濕(shi)試(shi)(shi)(shi)驗(yan)(yan)(yan)機進行試(shi)(shi)(shi)驗(yan)(yan)(yan)時,為了保證被(bei)試(shi)(shi)(shi)產(chan)(chan)品(pin)(pin)周圍氣氛能滿(man)足(zu)試(shi)(shi)(shi)驗(yan)(yan)(yan)規范所規定的試(shi)(shi)(shi)驗(yan)(yan)(yan)條件(jian)(jian),試(shi)(shi)(shi)驗(yan)(yan)(yan)箱工作(zuo)尺寸與被(bei)試(shi)(shi)(shi)驗(yan)(yan)(yan)產(chan)(chan)品(pin)(pin)外廓(kuo)尺寸之間應遵循以下幾點規則:
一、被試驗產品的(de)體積(D×W×H)不(bu)得(de)超過試驗箱有(you)效工作空間的(de)(20~35)%(推薦選用20%)。對于在(zai)試驗中(zhong)發熱的(de)產品推薦選用不(bu)大于10%.
二、被(bei)試(shi)(shi)驗(yan)產品的迎風斷面積與該斷面上試(shi)(shi)驗(yan)箱工作室總面積之比不大于(yu)(35~50)%(推薦選用35%)。
三、被試(shi)驗產品外(wai)廓表面距恒(heng)(heng)溫(wen)恒(heng)(heng)濕試(shi)驗箱壁的距離(li)至少(shao)保持100~150mm,(推(tui)薦選用150mm)。
以1立方米正方體(ti)箱子為例,面積比(bi)為1:(0.35~0.5)相當(dang)于體(ti)積之比(bi)為1:(0.207~0.354)。距箱壁100~150mm相當(dang)于體(ti)積之比(bi)為1:(0.343~0.512)。上述三(san)點(dian)規則彼(bi)此(ci)是(shi)相互(hu)依(yi)存和統(tong)一的(de)。選擇合適的(de)恒溫恒濕試(shi)驗機容(rong)積,不僅為企業降低資金(jin)運(yun)營(ying)成(cheng)本,增強設(she)備有效利用率,而(er)且試(shi)驗進行(xing)也可達到預期率。
針對產品大(da)小(xiao)、形狀等影響,不(bu)同(tong)(tong)產品選(xuan)擇內箱尺寸的(de)(de)大(da)小(xiao)也(ye)不(bu)盡相同(tong)(tong),恒(heng)溫恒(heng)濕試驗機容積(ji)大(da)小(xiao)的(de)(de)正確選(xuan)擇,對于購(gou)買方(fang)來(lai)說是(shi)一個十分(fen)重(zhong)要的(de)(de)決策。設(she)備購(gou)買小(xiao)了(le),試驗產品無法進入(ru)箱體或(huo)者影響產品試驗結果;設(she)備購(gou)買大(da)了(le),增加資金成本,減低設(she)備利用率。因此,恒(heng)溫恒(heng)濕試驗機容積(ji)大(da)小(xiao)的(de)(de)正確選(xuan)擇,是(shi)用戶必須掌握的(de)(de)一項知識。
出于以(yi)上幾點原(yuan)因(yin),有一(yi)下解決方案:
恒(heng)溫恒(heng)濕試(shi)驗(yan)箱的(de)(de)工(gong)(gong)作室容積至少應是被(bei)試(shi)驗(yan)產品外廓體(ti)積的(de)(de)3~5倍,恒(heng)工(gong)(gong)恒(heng)溫恒(heng)濕試(shi)驗(yan)機工(gong)(gong)作室內(nei)參(can)數〔如溫度(du)、濕度(du)、鹽(yan)霧(wu)沉(chen)降(jiang)率等〕的(de)(de)精度(du)指標都是在空載狀態下檢測的(de)(de)結(jie)果(guo),一旦置入被(bei)試(shi)驗(yan)件后,對(dui)試(shi)驗(yan)箱工(gong)(gong)作室內(nei)參(can)數的(de)(de)均勻(yun)性將產生影響,試(shi)驗(yan)件占(zhan)有的(de)(de)空間越大,這種(zhong)影響也就越嚴重。
被試(shi)(shi)驗(yan)(yan)件(jian)置入箱(xiang)體后擠(ji)占(zhan)了流(liu)暢(chang)的(de)通道,通道變窄將(jiang)(jiang)導致(zhi)氣(qi)流(liu)流(liu)速的(de)增加(jia)。加(jia)速氣(qi)流(liu)與(yu)被試(shi)(shi)驗(yan)(yan)件(jian)之間的(de)熱(re)交換(huan)。這與(yu)自然條件(jian)的(de)再現不(bu)符,在有關(guan)標(biao)(biao)準中(zhong)對涉及溫度試(shi)(shi)驗(yan)(yan)均規(gui)(gui)(gui)定(ding)(ding)試(shi)(shi)驗(yan)(yan)箱(xiang)內試(shi)(shi)驗(yan)(yan)樣件(jian)周圍的(de)空氣(qi)流(liu)速不(bu)應(ying)超過1.7m/s,以防(fang)止試(shi)(shi)驗(yan)(yan)樣件(jian)和(he)周圍氣(qi)氛產生不(bu)符合實際的(de)熱(re)傳導。在空載(zai)時試(shi)(shi)驗(yan)(yan)箱(xiang)內平(ping)均風(feng)(feng)速為(wei)0.6~0.8m/s(不(bu)超過1m/s),滿足(zu)規(gui)(gui)(gui)則(ze)1、2兩點要(yao)求所(suo)規(gui)(gui)(gui)定(ding)(ding)的(de)空間及面(mian)積比時,流(liu)場的(de)風(feng)(feng)速可能增大(50~100)%,平(ping)均最(zui)(zui)高風(feng)(feng)速為(wei)(1~1.7)m/s,滿足(zu)標(biao)(biao)準規(gui)(gui)(gui)定(ding)(ding)的(de)要(yao)求。如果在試(shi)(shi)驗(yan)(yan)中(zhong)不(bu)加(jia)限制地加(jia)大試(shi)(shi)驗(yan)(yan)件(jian)的(de)體積或迎風(feng)(feng)斷面(mian)積,則(ze)實際試(shi)(shi)驗(yan)(yan)時氣(qi)流(liu)風(feng)(feng)速將(jiang)(jiang)增大到(dao)超出試(shi)(shi)驗(yan)(yan)標(biao)(biao)準所(suo)規(gui)(gui)(gui)定(ding)(ding)的(de)最(zui)(zui)高風(feng)(feng)速,其試(shi)(shi)驗(yan)(yan)結果的(de)有效性將(jiang)(jiang)受到(dao)懷疑。
據實測試(shi)驗數據表明,流場(chang)中(zhong)迎風面(mian)與(yu)背風面(mian)的(de)(de)(de)(de)溫(wen)(wen)(wen)(wen)差(cha)可(ke)(ke)(ke)達到(dao)3~8℃,嚴(yan)重時可(ke)(ke)(ke)大到(dao)10℃以(yi)上(shang)(shang)。因此,必須盡量(liang)滿足規(gui)則1、2兩項要求,以(yi)保證被試(shi)驗產品(pin)周圍參數的(de)(de)(de)(de)均勻性。根據熱傳導的(de)(de)(de)(de)原(yuan)理,箱(xiang)(xiang)壁附近氣(qi)流的(de)(de)(de)(de)溫(wen)(wen)(wen)(wen)度(du)通常(chang)與(yu)流場(chang)中(zhong)心溫(wen)(wen)(wen)(wen)度(du)相(xiang)(xiang)差(cha)2~3℃,在高(gao)低(di)溫(wen)(wen)(wen)(wen)的(de)(de)(de)(de)上(shang)(shang)下限時,還可(ke)(ke)(ke)能達到(dao)5℃。箱(xiang)(xiang)壁的(de)(de)(de)(de)溫(wen)(wen)(wen)(wen)度(du)與(yu)箱(xiang)(xiang)壁附近流場(chang)的(de)(de)(de)(de)溫(wen)(wen)(wen)(wen)度(du)又(you)相(xiang)(xiang)差(cha)2~3℃(以(yi)箱(xiang)(xiang)壁的(de)(de)(de)(de)結(jie)構和材料而定(ding)),試(shi)驗溫(wen)(wen)(wen)(wen)度(du)與(yu)外(wai)界大氣(qi)相(xiang)(xiang)差(cha)越大,上(shang)(shang)述溫(wen)(wen)(wen)(wen)差(cha)也(ye)越大,因此,距(ju)(ju)箱(xiang)(xiang)壁(100~150mm)距(ju)(ju)離內的(de)(de)(de)(de)空間(jian)是(shi)不可(ke)(ke)(ke)利用空間(jian)。
恒溫恒濕試驗(yan)箱使用:待測品的(de)正確擺設方式(shi)?
恒(heng)溫恒(heng)濕試驗箱使用過程中(zhong),或多或少的(de)(de)會出現一(yi)些小故(gu)障、小難題,本文為(wei)您介紹(shao)的(de)(de)是:待(dai)測品的(de)(de)正確(que)擺(bai)設方式。希望能對您有所(suo)幫助(zhu),若(ruo)以下問(wen)題無法(fa)替您解決,請與我(wo)司技術部聯系(xi),我(wo)司將一(yi)一(yi)為(wei)您解決。
一、待測(ce)(ce)品的正確擺(bai)設方式:待測(ce)(ce)品放于恒溫恒濕箱(xiang)測(ce)(ce)試區當中,怎么擺(bai)設才算是正確的,規范是否(fou)有要(yao)求,與哪些因素有關?
分析正解:首(shou)先(xian)待(dai)測(ce)品(pin)應放(fang)置于箱體中中間,不能(neng)遮(zhe)住上面(mian)的(de)(de)風道(dao),也不能(neng)直(zhi)接(jie)放(fang)在箱體表面(mian),需(xu)要(yao)根據(ju)待(dai)測(ce)品(pin)的(de)(de)大小、形狀、體重、數量來(lai)調節樣品(pin)架的(de)(de)距離(li),如果需(xu)要(yao)外接(jie)電源則要(yao)將線(xian)(xian)統一接(jie)到(dao)測(ce)試孔旁邊,千萬不要(yao)讓線(xian)(xian)纏繞(rao)在一起,尤其(qi)要(yao)注意的(de)(de)是不要(yao)有(you)破損或其(qi)它(ta)危(wei)險(xian)因(yin)素的(de)(de)存在。
如果(guo)放到恒溫恒濕箱(xiang)的待測品是(shi)液體的一定(ding)要(yao)用容器(qi)盛裝(zhuang)好,如果(guo)是(shi)易揮發、易然易爆、有自然點的千萬不(bu)要(yao)設置(zhi)太高的溫度(du),也(ye)不(bu)要(yao)讓(rang)液體灑出來,如果(guo)是(shi)橡膠(jiao)(jiao)類的就(jiu)要(yao)用盤子墊著,因(yin)為橡膠(jiao)(jiao)一經高溫溶化可能會粘附到樣品架上(shang),清洗起來比較麻(ma)煩。
文章來源:東(dong)莞恒工設備有(you)限公(gong)司 業(ye)務聯系(xi):潘(pan)先(xian)生